The subject device was returned to olympus for investigation, but olympus could not reproduce the phenomenon.There was also no error code logged in the subject device.Olympus could not determine the cause of this phenomenon conclusively.The instruction manual of uhi-3 states that this phenomenon might occur by following factors.For those like children, whose abdominal cavity volume was small , insufflation at a high flow rate might cause the abdominal pressure to exceed the set pressure.Excessive gas was released from other gas supply devices like a laser device.So there was a possibility that the inappropriate handling by the facility and/or the patient's condition caused this phenomenon.Olympus also checked the manufacturing history of the subject device, and there was no irregularity found.The instruction manual of this device already mentions cautions for the device handling and the measures against excessive abdominal pressure.There were no further details provided.If significant additional information is received, this report will be supplemented.This report is being submitted as a medical device report in an abundance of caution.
|