While no serious injury resulted in this event, there has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21 cfr part 803.Functional quality testing was not performed since the attachment stopped working post production testing.The reported complaint was confirmed based on production testing.An actual temperature reading was not captured as the attachment stopped working post production testing.A root cause as to why this situation could have occurred could be determined based on quality observations.Bur end bearing failure, free roaming ball bearings and excessive debris most likely created friction within the head which resulted in temperature increase.All components looked dry and corroded with no evidence of lubrication.Additionally, the attachment failed all production requirements with the exception of water and chip air leak and illumination testing (sound testing was not performed).The last reported dentsply repair occurred on (b)(6) 2013 where the entire head assembly was replaced.
|