U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY PROFESSIONAL ESTYLUS 1:5 HIGH SPEED CONTRA ANGLE ATTACHMENT; HANDPIECE, AIR-POWERED, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY PROFESSIONAL ESTYLUS 1:5 HIGH SPEED CONTRA ANGLE ATTACHMENT; HANDPIECE, AIR-POWERED, DENTAL Back to Search Results
Catalog Number 774105
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
While no serious injury resulted in this event, there has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21 cfr part 803.Functional quality testing was not performed since the attachment stopped working post production testing.The reported complaint was confirmed based on production testing.An actual temperature reading was not captured as the attachment stopped working post production testing.A root cause as to why this situation could have occurred could be determined based on quality observations.Bur end bearing failure, free roaming ball bearings and excessive debris most likely created friction within the head which resulted in temperature increase.All components looked dry and corroded with no evidence of lubrication.Additionally, the attachment failed all production requirements with the exception of water and chip air leak and illumination testing (sound testing was not performed).The last reported dentsply repair occurred on (b)(6) 2013 where the entire head assembly was replaced.
 
Event Description
In this event a doctor reported that a estylus 1:5 attachment overheated and burned a patient.The patient was treated with oragel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESTYLUS 1:5 HIGH SPEED CONTRA ANGLE ATTACHMENT
Type of Device
HANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section D)
DENTSPLY PROFESSIONAL
901 west oakton st.
des plaines IL 60018
Manufacturer (Section G)
DENTSPLY PROFESSIONAL
901 west oakton st.
des plaines IL 60018
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key5223147
MDR Text Key31183482
Report Number1419322-2015-00122
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031145
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 10/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number774105
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/15/2015
Initial Date FDA Received11/13/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-