The patient underwent a generator replacement surgery on (b)(6) 2015 due to low battery, near end of service=yes.The explanted generator was received by the manufacturer for analysis.Analysis was completed.The reported end of service allegation, was duplicated in the analysis laboratory.However, the supply current tests did not meet functional specifications.These measurements demonstrate an increased current consumption for the device, potentially contributing to a premature end of life condition.The increased current consumption was isolated to a leaky capacitor (c6).The pulse generator performed according to functional specifications except for the tests that were due to the low battery condition.With the capacitor substitution for c6, the pulse generator module performed according to functional specifications.The cause for the c6 capacitors increase in leakage could not be determined.
|