(b)(4).Should the information be provided later, a supplemental medwatch will be sent.Batch #m91f20.The device was received with the blade broken off and not returned with the device; in addition the blade was discolored (blue) due to heat generated from contact between the blade and the tissue pad.This blade tip portion may have broken off of the device during transport to our analysis site.The reported complaint was for an activation issue.The device was functionally tested with a gen04.During functional testing an error code 5 was displayed.A probable cause of the device not activating and displaying an error code 5 is blade damage.Probable cause of blade discoloration is applying pressure between the instrument blade and tissue pad without having tissue between them.Once minor blade damage has occurred, subsequent activations may increase damage severity and result in an error code 5.The batch history records were reviewed with no anomalies noted during the manufacturing process.
|