The customer reported a positive result with a cyclesure 24 biological indicator (bi) after a completed sterrad cycle.The bi was incubated for 13 hours.The affected load was not released to patients.There was no report of infection, injury or harm to patient(s) associated with this issue.Although there is no report of patient injury or harm and no prior incidents have resulted in serious injury, advanced sterilization products (asp) has determined in this situation sterility cannot be assured.Therefore, as a matter of policy asp had decided to report all incidents of positive cyclesure 24 biological indicators.
|
(b)(4).Asp investigation summary: the investigation included a review of the device history record (dhr), trending of the product malfunction code, trending of lot number, system risk analysis (sra), and product retains analysis.The dhr was reviewed and the involved lot met manufacturer specifications at the time of release.No anomalies were observed that would contribute to the customer's experienced issue.Trending analysis for the product code of 'suspected positive bi" was reviewed from august 2014 through february 2015 and revealed a significant trend which was addressed through capa.Trending analysis by lot number was reviewed from 06/09/2015 to 10/23/2015 and trending was not exceeded.The sra indicates the risk associated with exposure to biohazardous, pathogenic or infectious material is "low." the suspect positive cyclesure® bi was not returned for evaluation.Thirty-two retains bis were subject to functional evaluation.All thirty-two bis met specification.Also, no damage or leakage was observed with the retains after processing.It is unlikely the suspected positive bi was caused by a manufacturing issue as the retains met functional specification, dhr review found no anomalies that would contribute to a positive bi result, and lot history review found trending was not exceeded in this lot.An issue with sterrad® performance is also unlikely as the cycle passed and the customer stated subsequent bis were negative for growth.The cyclesure® retains met functional specification.As a result, there is not an issue with product performance when used per the instructions for use (ifu).The issue will continue to be tracked and trended.
|