SYNTHES SELZACH DOUBLEAIRHOSE F/MINI AIR DRILL NO. 512.1; INSTRUMENT SURGICAL, ORTHO, PNEUMATIC, POWERED AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 512.150
Device Problem Air Leak (1008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Additional narrative: (b)(6).The 510k is not available as the device is currently in preamendment.The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device had a leaky hand piece and a leaking hose at the engine clutch.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was observed that the double air hose device had a leaky hand piece and a leaking hose at the engine clutch.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOUBLEAIRHOSE F/MINI AIR DRILL NO. 512.1
Type of Device
INSTRUMENT SURGICAL, ORTHO, PNEUMATIC, POWERED AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SYNTHES SELZACH
bohnackerweg 5
selzach CH254 5
SZ  CH2545
Manufacturer (Section G)
SYNTHES SELZACH
bohnackerweg 5
selzach CH254 5
SZ   CH2545
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5232499
MDR Text Key31748632
Report Number3000270450-2015-10232
Device Sequence Number1
Product Code HSZ
Combination Product (y/n)N
Reporter Country CodeSZ
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 11/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number512.150
Device Lot Number13820
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2015
Initial Date Manufacturer Received 11/10/2015
Initial Date FDA Received11/18/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/05/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-