Model Number SC-2366-50 |
Device Problems
Fracture (1260); High impedance (1291); Material Separation (1562)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 10/30/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
Additional suspect medical device components involved in the event: model #: sc-2366-50, serial #: (b)(4), description: linear 3-6 lead, 50cm.Model #: sc-8216-50, serial #: (b)(4), description: artisan surgical lead, 50cm.It is indicated that the devices will not be returned for evaluation; therefore a failure analysis of the complaint devices could not be completed.A review of the device history records will be conducted.If there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Event Description
|
A report was received that the patient was not receiving stimulation.High impedances were noted on the entire lead.X-ray showed lead fracture.No further course of action will be taken at this time.
|
|
Manufacturer Narrative
|
Additional information was received that the patient underwent a revision procedure wherein two leads was replaced.Device malfunction was suspected with the replaced leads.The patient was doing well postoperatively.
|
|
Event Description
|
A report was received that the patient was not receiving stimulation.High impedances were noted on the entire lead.X-ray showed lead fracture.No further course of action will be taken at this time.
|
|
Manufacturer Narrative
|
Sc-2366-50 (sn (b)(4)): device evaluation indicated that the complaint of high impedance was confirmed.Visual inspection revealed lead was distal end was completely separated from the lead body, just distal of electrode #8.The distal end section of the lead was not returned.It appeared that excessive tensile force was exerted onto the lead which resulted in the separation of the distal electrodes from the rest of the lead.Electrical testing could not be performed on the returned section of the lead.Sc-2366-50 (sn (b)(4)): device evaluation indicated that the complaint of high impedance was confirmed.Visual inspection revealed lead was distal end was completely separated from the lead body, just proximal of electrode #8.The distal end section (electrodes 1-7) of the lead was not returned.It appeared that excessive tensile force was exerted onto the lead which resulted in the separation of the distal electrodes from the rest of the lead.Electrical testing could not be performed on the returned section of the lead.
|
|
Event Description
|
A report was received that the patient was not receiving stimulation.High impedances were noted on the entire lead.X-ray showed lead fracture.No further course of action will be taken at this time.
|
|
Search Alerts/Recalls
|