• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Catalog Number 03666794689
Device Problems Thermal Decomposition of Device (1071); Corroded (1131)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/10/2015
Event Type  malfunction  
Manufacturer Narrative
Na.
 
Event Description
The caregiver called and alleged that the coaguchek xs meter (serial number (b)(4)) did not power on at all.She said she changed the batteries.When she changed them one of them felt hot and acid came out of one of the batteries.She advised that "one of the batteries seemed to be burned as it felt so hot to the touch".There was no injury from the battery.The suspect product was requested to be returned and replacement was sent.
 
Manufacturer Narrative
The meter was returned without batteries.New batteries were used but the device would not power on.There was a corroded battery contact.It was determined that due to the corroded contact the voltage supply was interrupted.The root cause for the heating and leakage of the battery was in defect in the battery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5245633
MDR Text Key31988803
Report Number1823260-2015-04623
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Reporter Occupation Other Caregivers
Type of Report Initial,Followup
Report Date 01/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Unknown
Device Catalogue Number03666794689
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/10/2015
Initial Date FDA Received11/24/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/05/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-