Catalog Number 03666794689 |
Device Problems
Thermal Decomposition of Device (1071); Corroded (1131)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/10/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Na.
|
|
Event Description
|
The caregiver called and alleged that the coaguchek xs meter (serial number (b)(4)) did not power on at all.She said she changed the batteries.When she changed them one of them felt hot and acid came out of one of the batteries.She advised that "one of the batteries seemed to be burned as it felt so hot to the touch".There was no injury from the battery.The suspect product was requested to be returned and replacement was sent.
|
|
Manufacturer Narrative
|
The meter was returned without batteries.New batteries were used but the device would not power on.There was a corroded battery contact.It was determined that due to the corroded contact the voltage supply was interrupted.The root cause for the heating and leakage of the battery was in defect in the battery.
|
|