Model Number 3788
Device Problem Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Type  Injury  
Manufacturer Narrative
(b)(4).This ipg serial number was included in a field correction.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
Device 1 of 2.Reference mfr.Report# 1627487-2015-20638.It was reported the patient ((b)(6)) experienced pocket heating while charging at the ipg site( date of event unknown).Consequently, a low energy replacement charger was provided to the patient.Implant date of the ipg is unknown at this time.On 08/01/2012, st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients.ࠁn increase in prior non-reported heating while charging events and other non-reported events was expected.
 
Event Description
Device 1 of 2.Reference mfr.Report# 1627487-2015-20638.Further follow up identified the issue of pocket heating while charging was resolved with charger replacement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key5245970
MDR Text Key32004663
Report Number1627487-2015-20637
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/26/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/28/2014
Device Model Number3788
Device Lot Number3710210
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/05/2015
Initial Date FDA Received11/24/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/15/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/25/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-001-C
Patient Sequence Number1
Patient Outcome(s) Other;
-
-