• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EON MINI CHARGING SYSTEM; SCS CHARGING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EON MINI CHARGING SYSTEM; SCS CHARGING SYSTEM Back to Search Results
Model Number 3721
Device Problem Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Corrective and preventive action (capa) investigation was performed.Pocket heating was confirmed. the investigation for capa (b)(4) associated with heating while charging (pocket heating) concluded that the charger was capable of transferring energy to the ipg at a rate that would cause heating of the ipg and/or charging wand of sufficient elevated temperature to cause pain and burns. the heating while charging was determined to be exacerbated by off-axis charging of shallow implanted ipgs and that all chargers were capable of elevated heating.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
Device 2 of 2.Reference mfr.Report# 1627487-2015-20637.
 
Manufacturer Narrative
This charger model was associated with a field correction.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
Device 2 of 2.Reference mfr.Report# 1627487-2015-20637.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI CHARGING SYSTEM
Type of Device
SCS CHARGING SYSTEM
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key5245974
MDR Text Key32004749
Report Number1627487-2015-20638
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/26/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/28/2013
Device Model Number3721
Device Lot Number3708638
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/05/2015
Initial Date FDA Received11/24/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/15/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/24/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-001-C
Patient Sequence Number1
Patient Outcome(s) Other;
-
-