|
Model Number 106 |
Device Problems
Improper or Incorrect Procedure or Method (2017); Computer Operating System Problem (2898)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/30/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that a patient was having a m106 generator implanted on (b)(6) 2015.The physician and company representative were having issues while trying to detect a heart rate.The pre-operative evaluation was not performed prior to surgery.The generator was already implanted as far laterally as possible, and no one was touching the generator while communicating with it.The wand was confirmed to have a functioning battery, and they changed the wand position as well.The impedance was within normal limits, and the output current was off.While trying to detect a heart rate, the programmer showed ???? for about 25 seconds.The sensitivity was changed to 3.At that time a heart rate of 85 bpm was seen briefly, but then went to ????.Sensitivities 2 and 4 were tested, but only ???? were seen.The generator was then replaced, and the issue resolved.Programming data from the surgery was received, and a heart rate of 85 bpm was detected at sensitivity 5 by the generator.This corresponds to the heart rate seen on the programming system during surgery.Interrogation was only performed at sensitivity level 5 while tachycardia detection was on.Therefore, no data is available regarding heart rate detection for sensitivity levels 1-4.The generator was received on (b)(6) 2015, but analysis has not been completed to date.
|
|
Event Description
|
Analysis was completed on the returned generator on 12/10/2015.The sense settings were evaluated with various loads, and various sense delay starts and sense drop outs were observed.Once the generator can was opened, possible contaminates were observed on the trimmed edge of the printed circuit board (pcb).Multiple measurements were not within specification, and the generator failed several electrical tests.After cleaning the trimmed edge of the pcb, the generator performed according to functional specifications.The generator was reassembled, and the sense settings were re-evaluated.There were no sense delays.The generator was instrumented in order to evaluate the delay in producing a heartbeat detection synch pulse.This synch pulse is received by the tablet and used to calculate the heartbeat rate.Test results of the synch pulse revealed a longer delay than what is expected.This delay may have been interpreted by clinicians, as non-detection of a heartbeat signal, causing them to attempt detection at a different setting or location on the patient.The removal of the tab from the pcb during the manufacturing process may have been the contributing factor for the undersensing.
|
|
Manufacturer Narrative
|
Inadvertently left off which test was out of specification.
|
|
Event Description
|
Testing was consistent with the known manufacturing error identified.The supply current was out of specification.
|
|
Search Alerts/Recalls
|
|
|