Section d4 other # field is populated with the di number.Additional suspect medical device components involved in the event: model #: sc-2218-70, serial #: (b)(4), description: linear st lead, 70cm, model #:sc-1110-02, serial #: (b)(4), description: precision implantable pulse generator (ipg), model #: sc-4316, lot #: 17711944, description: next generation anchor kit-sterile.
|
Device evaluation indicated that the source of the complaint was verified to be the associated leads whose cables were fractured at the clik anchor site, 1 cm from the set screw mark.Sc-1110-02/ sn: (b)(4) device evaluation indicated that the header was contaminated with blood, an impedance measurements were all fine.The ball seals were intact.The device passed the functional test and revealed no anomalies.Sc-2218-70/ sn: (b)(4) device evaluation indicated that five cables were fractured at the bent/kinked sections of the lead body at the clik anchor site.No exposed cables.Sc-2218-70/ sn: (b)(4) device evaluation indicated that four cables were fractured at the bent/kinked sections of the lead body at the clik anchor site.No exposed cables.Sc-4316/ sn: (b)(4) device evaluation indicated that the eyelets were torn, but there was no missing silicone material.
|