Model Number 3100A
Device Problems Device Maintenance Issue (1379); Noise, Audible (3273)
Patient Problem No Patient Involvement (2645)
Event Date 11/02/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The carefusion field service representative evaluated the unit and determined that the 3 ohm driver needed replacing as well as a 2,000 hour preventative maintenance performed.The field service representative performed the preventative maintenance, checked operations and the unit was meeting all specifications.The unit was placed back into service.(b)(4).
 
Event Description
The customer stated that the mean airway pressure does not go as far down as it needs to with the adjust knob.At a flow of 10 liters per minute the mean airway pressure will not go below 12 centimeters of water pressure.Also the oscillator is making a higher pitched sound compared to the other units.This issue was found during calibration and there was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
jill rittorno
22745 savi ranchpkwy.
yorba linda, CA 92887
7149229830
MDR Report Key5249903
MDR Text Key32235107
Report Number2021710-2015-02326
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/02/2015
Initial Date FDA Received11/25/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-