• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN, FORMERLY USSC PUERTO RICO INC ROTICULATOR ENDO GRASP* 5MM W/SPIN LOCK; MOTOR, SURGICAL INSTRUMENT, PNEUMATIC POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN, FORMERLY USSC PUERTO RICO INC ROTICULATOR ENDO GRASP* 5MM W/SPIN LOCK; MOTOR, SURGICAL INSTRUMENT, PNEUMATIC POWERED Back to Search Results
Model Number 174233
Device Problem Failure to Align (2522)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/18/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
According to the reporter, during a thoraco/wedge/segmental resection, a new sulu was opening, the jaws had been misaligned badly and did not engage properly.Stopped using the device and new one was opened to correct the problem.No patient involved.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
Post market vigilance (pmv) led an evaluation received one device opened by the account.This evaluation was based on a technical review of all data received from the site, a pmv review of manufacturing records, a pmv review of complaint trends, and an evaluation of the returned device.Engineering noted that the clevis curvature was observed bent or forced.This condition is not related to manufacturing process.A review of the device history record indicates this device lot number was released meeting all quality release specifications at the time of manufacture.Replication of damaged sheath condition may occur due to extreme handling during the application.The clevis curvature was observed bent or forced.Typically, this type of condition is observed in this unit when an excessive force or leverage is performed on the device during usage.As results of the excessive force the jaws will lose the force to open.It is possible to observe the jaws closed; however the jaws will be loose and will not remain closed with force.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
MOTOR, SURGICAL INSTRUMENT, PNEUMATIC POWERED
Manufacturer (Section D)
COVIDIEN, FORMERLY USSC PUERTO RICO INC
building 911-67
sabanetas industrial park
ponce PR 00731
Manufacturer (Section G)
COVIDIEN, FORMERLY USSC PUERTO RICO INC
building 911-67
sabanetas industrial park
ponce PR 00731
Manufacturer Contact
sharon murphy
60 middletown ave
north haven, CT 06473
2034925267
MDR Report Key5250031
MDR Text Key32284644
Report Number2647580-2015-00831
Device Sequence Number1
Product Code GET
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K914753
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2020
Device Model Number174233
Device Catalogue Number174233
Device Lot NumberP5F0138X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/20/2015
Initial Date FDA Received11/25/2015
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received12/30/2015
06/15/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2015
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-