COVIDIEN, FORMERLY USSC PUERTO RICO INC ROTICULATOR ENDO GRASP* 5MM W/SPIN LOCK; MOTOR, SURGICAL INSTRUMENT, PNEUMATIC POWERED
|
Back to Search Results |
|
Model Number 174233 |
Device Problem
Failure to Align (2522)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/18/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
According to the reporter, during a thoraco/wedge/segmental resection, a new sulu was opening, the jaws had been misaligned badly and did not engage properly.Stopped using the device and new one was opened to correct the problem.No patient involved.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Manufacturer Narrative
|
Post market vigilance (pmv) led an evaluation received one device opened by the account.This evaluation was based on a technical review of all data received from the site, a pmv review of manufacturing records, a pmv review of complaint trends, and an evaluation of the returned device.Engineering noted that the clevis curvature was observed bent or forced.This condition is not related to manufacturing process.A review of the device history record indicates this device lot number was released meeting all quality release specifications at the time of manufacture.Replication of damaged sheath condition may occur due to extreme handling during the application.The clevis curvature was observed bent or forced.Typically, this type of condition is observed in this unit when an excessive force or leverage is performed on the device during usage.As results of the excessive force the jaws will lose the force to open.It is possible to observe the jaws closed; however the jaws will be loose and will not remain closed with force.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.
|
|
Search Alerts/Recalls
|
|
|