The device referenced in this report has not yet been returned to olympus for evaluation.The exact cause of the reported phenomenon could not be determined at this time.If additional information or if device is received at a later date, this report will be supplemented.The instruction for use states, always operate the forceps lightly- excessive force on the finger grip (either pushing or pulling) can cause deterioration and breakage.
|
Olympus was informed that during an unspecified procedure, the tip of the device broke off and fell inside the patient while the physician was trying to remove an essure coil that did not deploy properly.There was minor bleeding that occurred as a result of removing the coil.The device fragment was not found during straining of the patient's cavity after the procedure.In addition, the patient required a longer stay and was admitted for a 24 hour observation and will be requiring additional procedure to remove the coil.It is unknown at this time if the broken piece was retrieved from the patient.No further information was provided.
|