Model Number NGP680300
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).The field service specialist (fss) ordered a new i/v pole bottle hanger to the site to be fitted by the local amo representative during a preventative maintenance site visit.All pertinent information available to abbott medical optics has been submitted.
 
Event Description
The surgery center reported that the i/v pole bottle hanger was broken and it was found to have the fulcrum pin missing with the whitestar signature system.Additional communication on (b)(6) 2015 verified that the date of the event was unknown because the machine was moved from another location and during installation at its current site, the fss found the problem because tape had been placed on the machine to hold the bottle hanger in position.There was no patient involvement reported and no patient treatments delayed or cancelled.
 
Manufacturer Narrative
Device manufacture date: 01/21/2010.All pertinent information available to abbott medical optics has been submitted.
 
Manufacturer Narrative
A record review was performed.A product deficiency review was performed and there is no product deficiency identified.A document and trending was reviewed.There is not a recognizable adverse trend.The risks and mitigations associated with the complaint issue are identified in existing risk documents and no new risks were identified as part of this investigation.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WHITESTAR SIGNATURE SYSTEM
Type of Device
PHACOFRAGMENTATION UNIT
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
Manufacturer (Section G)
ABBOTT MEDICAL OPTICS INC.
510 cottonwood drive
milpitas CA 95035
Manufacturer Contact
valerie sedzicki
1700 east st. andrew place
santa ana, CA 92705
7142478567
MDR Report Key5251927
MDR Text Key32199487
Report Number3006695864-2015-00878
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K060366
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNGP680300
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/02/2015
Initial Date FDA Received11/25/2015
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received12/10/2015
01/28/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/21/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-