MAUDE Adverse Event Report: STRYKER TRAUMA KIEL LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X45 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X45 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 18965045S
Device Problems Break (1069); Failure to Osseointegrate (1863)
Patient Problem Inadequate Osseointegration (2646)
Event Date 11/09/2015
Event Type  Injury  
Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event Description
The sales rep has reported on behalf of the customer that a patient has undergone revision surgery on a t2 ankle arthrodesis nail, due to alleged lack of fusion.The sales rep has reported that when the patient was opened up it was noticed that two screws had snapped.The surgery was completed successfully by inserting two new screws and a the planned longer nail.
 
Manufacturer Narrative
The two broken locking screws were classified as primary product during investigation.No deviations were found during review of the manufacturing and inspection documents (dhr).The locking screws returned were documented as faultless prior to distribution.During investigation no material, design or manufacturing related issues were found.The breakage surfaces of both screws are smooth and show lines of rest; both screws broke due to a fatigue fracture.The windings of the screw threads are partly deformed / abraded and destroyed, most likely due to heavy postoperatively contact with the nail hole rims, combined with nail hole rim contact during screw implantation.The screws broke due to a postoperatively overload, the implants were loaded several times with too high bending forces.Due to this the screws cracked and broke step by step till they got completely broken after approx.Eight months.The patient¿s height and weight were not provided, obesity cannot be excluded.The customer reported that the nail was exchanged to a longer nail due to ¿lack of fusion¿ and to ¿prevent the nail migrating medically in the tibia as it had done with 150 nail.¿ a lack of fusion could not be confirmed or denied; a fusion or non-fusion could not be determined based on the provided x-ray.The position of the inserted compression screw and the contact-marks on the partly threaded screw indicates that the tibia-talar-compression was performed; therefore it is most likely that the compression of the tibia-talar-joint was successfully done.If the talar-calcaneal-compression was also successfully done could not be determined due to missing information, but the fact that the screws within the calcaneal bone got broken indicates that most likely the joint was still free or partly moveable over the whole implantation time of approx.8 months.An inadequate compression of this joint most likely happened.A medically migration was not visible on the provided x-ray; therefore the migration could not be confirmed or denied.The ifu includes several adverse effects (obesity, intra-operatively implant damages, post-operatively loads, etc.) that can lead to implant failures like screw breakages.The successfully joint compression of both joints is described in detail in the operative technique.As a secondary issue it was found that the distal nail holes were drill marked; the most distal hole was completely missed and the outer nail surface was damaged by the drill on the medial side.It is likely that the target device was pre-damaged and / or the nail holding screw was not tightened to the nail and / or bending forces were applied to the drill / target device during drilling.The ifu and operative technique includes that every instrument shall be checked prior to every surgery regarding its functionality.Based on the investigation and all given information the screw breakages are not manufacturer related; they broke due to postoperative overloading (patient related) combined with damages during implantation and most likely with an inadequate compressed subtalar joint (user related).No non-conformity identified.
 
Event Description
The sales rep has reported on behalf of the customer that a patient has undergone revision surgery on a t2 ankle arthrodesis nail, due to alleged lack of fusion.The sales rep has reported that when the patient was opened up it was noticed that two screws had snapped.The surgery was completed successfully by inserting two new screws and the planned longer nail.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X45 MM
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5254959
MDR Text Key32312437
Report Number0009610622-2015-00565
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeGB
PMA/PMN Number
K032244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/30/2019
Device Catalogue Number18965045S
Device Lot NumberK025CBE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/09/2015
Initial Date FDA Received11/30/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/08/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age65 YR
-
-