• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION PRECISION SPECTRA®; SPINAL CORD STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION PRECISION SPECTRA®; SPINAL CORD STIMULATOR Back to Search Results
Model Number SC-2316-50
Device Problems Bent (1059); Fracture (1260); High impedance (1291); Kinked (1339)
Patient Problem Inadequate Pain Relief (2388)
Event Date 11/06/2015
Event Type  Injury  
Manufacturer Narrative
Additional suspect medical device component involved in the event: model #: sc-2316-50, serial #: (b)(4), description: infinion 1x16 perc lead kit-50 cm.Sc-2316-50, (b)(4).The source of the complaint was verified to be associated with infinion leads whose cables were fractured.All cables were fractured at the bent/kinked sections of the lead body, at the clik anchor site, 1 cm from the set screw mark.Fractured location was about 8 inches from the distal tip.No cables were exposed.
 
Event Description
A report was received that during the device analysis, it was determined that the leads were fractured at the bent/kinked sections of the lead body.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION SPECTRA®
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key5256310
MDR Text Key32339940
Report Number3006630150-2015-03090
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/31/2015
Device Model NumberSC-2316-50
Other Device ID NumberM365SC2316500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/06/2015
Initial Date FDA Received11/30/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/02/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age51 YR
-
-