The device was received for evaluation.Condition upon receipt: un-sealed sample, part number 1885078hse, from lot number 0207851669 received.Equipment used: microscope (zeiss stemi 2000c between 0, 65 to 5, 0 magnification settings), calipers.Evaluation: when compared to the assembly drawing, visually, the inner assembly was stretched out of its support area by approximately 0.22¿ which would have resulted in the reported event.The inner assembly spins however it is with resistance, which is likely caused by the stretching (deformation) of the spiral wrap.There was no portion of the device that became detached.When viewed under magnification, there was no additional evidence to isolate a cause therefore manufacturing, supplier, and misuse / mishandling cannot be ruled out as potential causes.Methods - actual device evaluated.(b)(4).
|