There are no reported injuries associated with this event.The reported symptom relating to during use, the process chamber will fill with water but does not go any further than the wash / disinfect cycle was attributed to the basin drain cup, (p/n 11013); it's positioning could have contributed to an air pocket that may not allow the liquid to be properly be detected resulting in a safety float error.The drain cup was re-aligned and the system performed as intended.The reported issue relating to 783 wash only (f6) cycles being performed since (b)(6) 2014, cannot be clearly determined as follow up calls have gone unanswered; however, it is attributed to user error in not following the intended use of the device.Custom ultrasonics is reporting this event out of an abundance of caution.
|
On (b)(6) 2015, the customer called requesting service due to during use, the process chamber will fill with water but does not go any further than the wash / disinfect cycle.Scopes being processed at the time were processed alternatively.On-site inspection on (b)(6) 2015 by a custom ultrasonics service technicians found after inspection of the system log, the customer may have performed wash and rinse cycles only during processing of scopes.There has been no report of patient injury; however, the customer has not been responsive to inquiries.
|