• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND GMBH SORIN S5 SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND GMBH SORIN S5 SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number UNKNOWN
Device Problem Failure to Auto Stop (2938)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/04/2015
Event Type  malfunction  
Manufacturer Narrative
Part number and serial number not provided yet.This information will be provided in a supplemental report if and when made available.The manufacturing date is unknown because the serial number has not been provided.Sorin group (b)(4) manufactures the sorin s5 system.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the level detection system for the sorin s5 system failed to alarm and did not stop the pump during a procedure.There was no report of patient injury.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that the level detection system for the sorin s5 system failed to alarm and did not stop the pump during a procedure.There was no report of patient injury.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the sorin s5 system.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the level detection system for the sorin s5 system failed to alarm and did not stop the pump during a procedure.There was no report of patient injury.Photos of the set-up being used were taken and sent to sorin group (b)(4) for further investigation.Evaluation of the customer photos concluded that the customer was not using the level pad according to the instructions for use.This issue was caused by user error.As a serial number and part number were not provided by the facility, a dhr review could not be performed.Sorin group (b)(4) will continue to monitor the market for trends related to this type of issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN S5 SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND GMBH
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
carrie wood
14401 w. 65th way
arvada, CO 80004
3034676461
MDR Report Key5262918
MDR Text Key32578978
Report Number9611109-2015-00597
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUNKNOWN
Device Catalogue NumberN/A
Device Lot NumberN/A
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/05/2015
Initial Date FDA Received12/03/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/03/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-