MAUDE Adverse Event Report: CONMED LINVATEC FLX-LITE, STERILE, SINGLE USE, FLEXIBLE SURGICAL; LIGHT, EXAMINATION, MEDICAL, BATTERY POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONMED LINVATEC FLX-LITE, STERILE, SINGLE USE, FLEXIBLE SURGICAL; LIGHT, EXAMINATION, MEDICAL, BATTERY POWERED Back to Search Results
Catalog Number 3034A
Device Problems Sparking (2595); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  No Answer Provided  
Manufacturer Narrative
As of this filing, the damaged examination light has not yet been returned/received from the user facility for evaluation.The investigation remains in process.A supplemental and final report will be filed upon the completion of the investigation.
 
Event Description
The customer reported that while preparing to intubate a patient and getting the flex-lite, sterile prepared, the light bulb popped and sparks came out for a few seconds when the light bulb was activated.The procedure was completed successfully with no delay and no injury.
 
Manufacturer Narrative
Conmed has been advised by the distributor that the device has been lost and will not be returned.Without the actual product, an evaluation could not be performed and the root cause of the alleged problem was unable to be determined.A review of the device history record showed that this lot was manufactured on 16-jan-2014 in a lot of (b)(4) units.There were no discrepancies noted during the manufacturing process that could have caused or contributed to the reported breakage.There have been no other similar complaints received for this item and lot number combination.A two-year review of complaint history shows no other similar complaints for this item.A supplier corrective action has been generated to address the reported issue.Additionally, the supplier is updating the dfmea to address the risk of the bulb popping upon activation.According to the supplier, there has been no previous report for this device failure mode.The instructions for use (ifu) provides the following contraindications and cautions: - this device is contraindicated for use in the presence of flammable gasses or other materials capable of ignition.- do not use if package is damaged or opened.- avoid having bulb make contact with hard objects.Distributor lost device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLX-LITE, STERILE, SINGLE USE, FLEXIBLE SURGICAL
Type of Device
LIGHT, EXAMINATION, MEDICAL, BATTERY POWERED
Manufacturer (Section D)
CONMED LINVATEC
11311 concept blvd.
largo FL 33773
Manufacturer (Section G)
CONMED LINVATEC
11311 concept blvd.
largo FL 33773
Manufacturer Contact
brenda johnson
11311 concept blvd.
largo, FL 33773
7273995515
MDR Report Key5262926
MDR Text Key32888832
Report Number1017294-2015-00067
Device Sequence Number1
Product Code KYT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/30/2018
Device Catalogue Number3034A
Device Lot Number0114C
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/05/2015
Initial Date FDA Received12/03/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/05/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/16/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-