|
Model Number 31180698E |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/30/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
This device is used for therapeutic purposes.Concomitant medical products: bur 31180698e visao cvd diamd.6x98mm l, 510k: exempt , lot# 0209568969, mfr date:2015-05-06, use before: 2019-05-05.(b)(4).The products have not been returned for analysis.
|
|
Event Description
|
It was reported that the tip of the bur broke off and was lodged in patient during cochlear implant.It was removed via suction, without complications.The patient was fine when woken from procedure.
|
|
Manufacturer Narrative
|
Devices received on (b)(4) 2015.Devices evaluated on (b)(4) 2015.During product analysis, the devices were compared to the assembly drawing.When lot # 0209674535 was viewed under magnification, the distal tip was missing and not returned.The shaft broke just inside the outer tube meaning the piece that broke off measured approximately (0.12¿).The fracture was cone shaped which suggests a compression fracture however this cannot be confirmed.Under magnification there was no additional evidence to isolate a cause regarding lot # 0209568969, the tip was present and measured 0.64mm which is not consistent with the reported event.The inner assembly had 0.15¿ axial play which exceeds the allowed tip tolerance of +-0.015¿.The overall length is a reference dimension on the drawing (3.814¿) which carries no tolerance requirement; the overall length was measured at approximately 3.82¿ which is within a reasonable range.Functionally the bur would load into a handpiece, run at 80,000 rpm, and cut saw bone with no issues.The excess axial play may have been the result of extended use and wear or may have been an out of specification condition from manufacturing.(b)(4).
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|