ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/19/2015
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when addition reportable information becomes available.(b)(4).
 
Event Description
A customer reported the unit was locking when stepping on the foot pedal during a vitrectomy procedure.The event did not prevent the conclusion of the procedure.There was no harm to the patient.Additional information has been requested.
 
Manufacturer Narrative
No further information was able to be obtained from this customer.With no additional, related information provided, the customers reported event was not able to be confirmed.A review of the customer¿s complaint history for the last 24 months did not show any previous complaints of this kind against the system.The system was manufactured on november 20, 2010.Based on qa assessment, the product met specifications at the time of release.The footswitch was manufactured on october 18, 2010.Based on qa assessment, the product met specifications at the time of release.Footswitch was received for evaluation.An assessment of the returned sample revealed the left and right heel switches would stick when pressed.The left heel switch was opened up and dried balanced saline solution (bss) ingress was seen on the pin and pin hole.The right heel switch was unable to be opened due to a stuck screw.Evidence of bss ingress indicates the same cause as the left heel switch.There was no information on the system.The root cause of the reported event can be attributed to bss ingress on the pins of the heel switches, causing them to stick.An internal investigation has been opened to address the bss issue.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The system was examined and the reported event was replicated.The footswitch was replaced to resolve the issue.The system was tested and found to meet product specifications.A review of the customer¿s complaint history for the last 24 months did not show any previous complaints of this kind against the system.The system was manufactured on november 20, 2010.Based on qa assessment, the product met specifications at the time of release.The footswitch was manufactured on october 18, 2010.Based on qa assessment, the product met specifications at the time of release.The system was found to meet specifications.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
eddie darton, md, jd
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175686660
MDR Report Key5267434
MDR Text Key32883045
Report Number2028159-2015-09802
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 06/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/19/2020
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Other Device ID Number3.00.61
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/19/2015
Initial Date FDA Received12/04/2015
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received07/01/2016
06/14/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/20/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-