Model Number 3799 |
Device Problems
Low impedance (2285); Temperature Problem (3022)
|
Patient Problems
Pain (1994); Burning Sensation (2146)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Event Description
|
It was reported the patient ((b)(6)) experienced burning/heating sensation with stimulation (date of event unknown).In turn the patient stopped using stimulation.Additionally, the patient experienced pain while sitting/ putting pressure at the ipg site.System diagnostics determined low impedances on the system.Surgical intervention was taken where the ip g was explanted and replaced which resolved the issue.
|
|
Manufacturer Narrative
|
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|