• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Intermittent Continuity (1121); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/11/2015
Event Type  malfunction  
Event Description
On (b)(6) 2015 it was reported that the physician's tablet was having intermittent communication issues.The physician was not able to use the system at all on the patient in the office at that time however the therapeutic consultant was able to use his dosing system on the patient with no issues.The serial cable on the tablet was removed and then reinserted with no resolution of the communication issues.The 9v battery was checked and confirmed to remain on for >25 seconds.The tablet was not plugged into the wall during use.The therapeutic consultant confirmed that when his wand and serial cable were substituted on the tablet, he was getting intermittent issues as well.When he used the physician's wand and serial cable with his tablet he was able to successfully communicate with his demo product after an initial communication error.The tablet was returned for product analysis on (b)(6) 2015.Product analysis is still underway and has not yet been completed.
 
Event Description
Product analysis was completed on the tablet on 12/16/2015.No anomalies associated with the tablet performance were noted during testing using the ac adapter or the main battery with a full charge.The tablet performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5270935
MDR Text Key33293279
Report Number1644487-2015-06660
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 11/11/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11.0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2015
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 11/11/2015
Initial Date FDA Received12/07/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/27/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-