• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION VELA VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION VELA VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number VELA COMP D
Device Problem Delamination (2904)
Patient Problem No Patient Involvement (2645)
Event Date 11/09/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).In the event the device is received for evaluation or additional information is received a follow-up report will be submitted.(b)(4).
 
Event Description
The customer stated that the front panel has delamination and the touchscreen is unresponsive.This incident was found upon testing therefore no patient involvement.
 
Manufacturer Narrative
Results of investigation: the carefusion failure analysis lab evaluated the suspect faulty front panel and was able to reproduce the reported event of there being liquid ingress and an inactive touch screen.This issue is part of an internal investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELA VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
CAREFUSION
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
jill rittorno
22745 savi ranch pkwy
yorba linda, CA 92887
7149229830
MDR Report Key5271835
MDR Text Key33075198
Report Number2021710-2015-02382
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K093094
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVELA COMP D
Device Catalogue Number16532-02
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/09/2015
Initial Date FDA Received12/07/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/11/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-