Model Number 5392 |
Device Problems
Nonstandard Device (1420); Device Displays Incorrect Message (2591)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/13/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product problem: analysis confirmed the customer comment of a generated error via the error log and attributed it to the main printed circuit board being out of specification in an electrical manner.It was also noted that the keyboard was scratched.All found defective parts were replaced and all other identified issues were resolved.(b)(4).
|
|
Event Description
|
It was reported that while in use on a pacer-dependent patient the external pulse generator generated an "error, device needs to be restarted" message.It was noted that there was no affiliated error number with the message.It was also noted that the heart rate was still showing spikes on the patient monitor when this occurred.The generator was changed out immediately when the error was noticed, and it was not noted what all of the settings had been.The generator was returned for service.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Further analysis was performed on the main printed circuit board assembly.Visual inspection revealed no anomalies.Bench analysis found that the device powered up normally and passed all functional tests.While the device was running an error appeared on the display, the error appeared more often when the device was placed in an oven at 70 degrees celsius.The device did not fail at cold temp.The die stack was replaced, the device ran for 12 hours in the oven with no failures.The error was documented in the error log.Conclusion: the reported event was confirmed, the die stack was defective.This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.
|
|
Search Alerts/Recalls
|