• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. ATTUNE SPACER BLOCK; KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. ATTUNE SPACER BLOCK; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 254401014
Device Problems Break (1069); Device Or Device Fragments Location Unknown (2590)
Patient Problem No Information (3190)
Event Date 11/26/2015
Event Type  malfunction  
Manufacturer Narrative
This complaint is under investigation.Depuy will notify the fda of the results of the investigation once it has been completed.Hat was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
One of the location lugs has a small piece that has broken off.Broken piece not returned to engineering.Item was returned in kit from hospital no other information was provided.This case has been reported by the loan kit technician during loan kit inspection.
 
Manufacturer Narrative
This complaint remains under investigation.Depuy will notify the fda of the results of the investigation once it has been completed.
 
Manufacturer Narrative
Conclusion and justification status: the complaint states one of the location lugs has a small piece that has broken off.The investigation confirmed the failure mode as reported.Care needs to be used when assembling these connections, as damage may occur resulting in the type of failure seen here.However, there is insufficient information supplied with this complaint to determine whether sufficient care was or was not taken.It should be noted that (b)(4) was released on 08 jul 2014 in which the damage to the balseals was evaluated.The pra concluded that there is no patient harm as none of the failure modes have led to any patient harm.Furthermore, the assembly and disassembly of this device is carried out by the scrub nurse on a table away from the joint space, additionally the device cannot be disassembled within the joint space.It should also be noted that this issue will be further monitored in post market surveillance the complaint shall be closed with an undetermined conclusion and entered into the complaints system and monitored through trend analysis.If further information is received the complaint shall be reopened and investigated further.Depuy still considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE SPACER BLOCK
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key5285857
MDR Text Key33169579
Report Number1818910-2015-37186
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 01/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254401014
Device Lot NumberBFA0P43
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/30/2015
Initial Date FDA Received12/10/2015
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received01/11/2016
01/27/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/09/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-