• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problems Cut In Material (2454); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/26/2015
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of monitor sn (b)(4) has been completed.The reported problem (constant ongoing, exposed wires) was confirmed.As received, the monitor failed incoming testing.Upon evaluation, the belt receptacle was pulled from the monitor case, damaging the internal pulse wire.The cause of the constant ongoing and exposed wires is the damaged receptacle and wires.The root cause of the damaged receptacle and wire is excessive force placed on the monitor receptacle while an electrode belt was attached.No adverse event resulted from the damaged receptacle.
 
Event Description
A us distributor contacted zoll to report that a patient's device was constantly producing gong alerts and there were wires exposed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
katelynn mains
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key5287180
MDR Text Key33749604
Report Number3008642652-2015-07884
Device Sequence Number1
Product Code MVK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,study,user facili
Reporter Occupation Other
Type of Report Initial
Report Date 12/04/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2015
Initial Date Manufacturer Received 11/20/2015
Initial Date FDA Received12/10/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-