Visual assessment of the device confirmed the reported breakage.The threaded distal tip has broken off at the transition of the threads to the.110 diameter of the shaft.The break area shows no material voids.Material condition was inspected and found to meet print specification.The proximal end of the handle is damaged, likely the result of impact from a mallet during use.This method of use is in conflict to the device ifu for both the anchor and awl-dilator.The healicoil ifu states ¿awl dilator: rotate the handle of the awl dilator clockwise until the laser mark is flush with the surface of the bone.To remove the awl dilator from the insertion site, rotate the device counterclockwise¿.Per the awl dilator ifu under precautions ¿as with any surgical instrument, careful attention should be exercised to ensure that excessive force is not placed on the instrument.Excessive force can result in instrument failure¿.After the evaluation, the root cause for the reported issue was determined to be user error.A review of the device history record was performed which confirmed no inconsistencies.
|