Model Number 840
Device Problem Poor Quality Image (1408)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/06/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The service engineer (se) evaluated the ventilator, and verified the customer reported malfunction.The se determined that the graphic user interface (gui) printed circuit board (pcb) needs to be replaced.However, the repair has not been completed.
 
Event Description
It was reported that, during patient use, the ventilator lower display became blurred and difficult to read.However, ventilation was not affected.It is unknown if the patient was transferred to another device.There is no report of death or serious injury associated with this event.
 
Manufacturer Narrative
The device was repaired and the reported issue was isolated to interface between the device and the failed component.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Type of Device
CONTINUOUS VENTILATOR
Manufacturer (Section D)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI 
Manufacturer (Section G)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI  
Manufacturer Contact
thom mcnamara
15 hampshire street
mansfield, MA 02048
5084524811
MDR Report Key5292995
MDR Text Key34004016
Report Number8020893-2015-01715
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K970460
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 12/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number840
Device Catalogue Number840
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/20/2015
Initial Date FDA Received12/14/2015
Supplement Dates Manufacturer Received11/20/2015
Supplement Dates FDA Received12/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/14/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-