Model Number 840 |
Device Problem
Poor Quality Image (1408)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/06/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The service engineer (se) evaluated the ventilator, and verified the customer reported malfunction.The se determined that the graphic user interface (gui) printed circuit board (pcb) needs to be replaced.However, the repair has not been completed.
|
|
Event Description
|
It was reported that, during patient use, the ventilator lower display became blurred and difficult to read.However, ventilation was not affected.It is unknown if the patient was transferred to another device.There is no report of death or serious injury associated with this event.
|
|
Manufacturer Narrative
|
The device was repaired and the reported issue was isolated to interface between the device and the failed component.If information is provided in the future, a supplemental report will be issued.
|
|