DEPUY SYNTHES POWER TOOLS BLACKMAX-NEURO LOW PRESSURE; MOTOR, DRILL, PNEUMATIC - HANDPIECE
|
Back to Search Results |
|
Catalog Number BLACKMAX-N-LP |
Device Problems
Break (1069); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Initial reporter's phone number: (b)(6).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and found that the handpiece device had no rotation and was not running due to broken vanes in the motor.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to the worn out motor.This is further defined as wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported by the (b)(6) that during service and evaluation, it was observed that the motor device had no rotation and was not running due to broken vanes in the motor.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|