EDWARDS LIFESCIENCES, PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER WITH AMC THROMBOSHIELD; CATHETER, OXIMETER, FIBEROPTIC Back to Search Results
Model Number 774HF75
Device Problem Defective Component (2292)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/23/2015
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for evaluation; it was discarded at the hospital.Without return of the unit it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.It is not known if some procedural factors may have contributed to the event.No corrective actions will be taken at this time.Lot number was not provided; therefore review of the manufacturing records could not be completed.
 
Event Description
As reported, in this swan ganz catheter the balloon could be inflated in the off position.On the swan ganz catheter there is a volume limited syringe which has an on/off switch located at its base, which is designed to prevent accidental inflation; however it did not function.Balloon inflation was tested prior to use.The issue was noticed when teaching a nurse how to wedge the device.The catheter was used with extra vigilance after wedging it.There was no allegation of patient injury.The device was discarded at the hospital.
 
Manufacturer Narrative
A review of the manufacturing records indicated that the product met specifications upon release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER WITH AMC THROMBOSHIELD
Type of Device
CATHETER, OXIMETER, FIBEROPTIC
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
lynn selawski
1 edwards way
irvine, CA 92614
9497564386
MDR Report Key5306868
MDR Text Key33784418
Report Number2015691-2015-03416
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 11/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/10/2017
Device Model Number774HF75
Device Lot Number60173678
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/24/2015
Initial Date FDA Received12/17/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/11/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/10/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-