This device is used for therapeutic purposes.(b)(4).The device was returned for evaluation.Condition upon receipt: 1 un-sealed sample, part number 1883670hs, from lot number 62671600 received; there was evidence of biological contaminants [based off of the reactivity with hydrogen peroxide].Equipment used: microscope (zeiss stemi 2000c between 0, 65 to 5, 0 magnification settings), m4 handpiece (part 1898200t), calipers.Evaluation: when compared to the assembly drawing and inner assembly drawing of the distal portion of the inner assembly became detached which would have resulted in the reported event.The break point corresponds to the proximal side of the outer tube radius.Included with the returned device were 2 spiral wrap fragments measuring less than 1/8¿ each.When viewed under magnification, the spiral wrap was coiled in on itself in a clockwise direction which likely indicates the break occurred while being used in forward direction.There was gouging in the support area just proximal to the distal tip (both the inside and outside diameters) which indicates excess pressure was being applied during use.There was damage to the hubs that is consistent with improper loading: dimples on the front hub prior to the locking area caused by the handpiece locking mechanism; and damage to the inner hub chevrons caused by the handpiece drive mechanism.The instructions for use has detailed instructions for properly loading a bur/blade into the handpiece and warn that excessive pressure applied to a bur/blade may cause a fracture.Note: [excess: exceeded sufficient pressure required for operation].Functionally, the bur loaded securely into a handpiece.There was no evidence of improper manufacturing and therefore has been ruled out as a likely cause.
|