The customer complained of erroneous results for 2 patients tested for free thyroxine (ft4 ii), ft3 - free triiodothyronine (ft3 iii), and thyrotropin (tsh).It is not known if erroneous results were reported outside of the laboratory.This medwatch will cover ft4 ii.Refer to medwatch with (b)(6) for information on the ft3 iii erroneous results and medwatch with (b)(6) for information on the tsh erroneous results.On (b)(6) 2015 patient 1 initial tsh result was approximately 1.2 uui/ml, the ft4 ii result was over 2 times the upper reference range (the specific result was not provided), and the ft3 iii result was over 2 times the upper reference range (the specific result was not provided).A new sample was obtained on (b)(6) 2015 and the initial tsh result was approximately 1.2 uui/ml, the ft4 ii result was over 2 times the upper reference range (the specific result was not provided), and the ft3 iii result was over 2 times the upper reference range (the specific result was not provided).Both samples were repeated on an abbott instrument where the tsh result was in the normal range and the ft4 ii and ft3 iii results were just below the upper reference range.There is a suspicion that this patient sample contains a resistance for thyroid hormone.Patient 2 initial tsh result was 0.64 uui/ml, the ft4 ii result was 77 pg/ml and the ft3 iii result was 9.2 pg/ml.The sample was repeated on an abbott instrument and the tsh, ft4 ii and ft3 iii results were in the normal ranges.It is not known if either patient was adversely affected.No adverse event was reported.The e601 analyzer serial number was (b)(4).
|