Model Number M00556581
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/02/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4) guidewire distal tip detached, exposing the tip of the metal corewire.The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that a jagwire guidewire was used during an endoscopic retrograde cholangiopancreatography procedure performed on (b)(6) 2015.According to the complainant, during introduction, the hydrophilic tip became detached, exposing the tip of the metal corewire.No part of the guidewire detached inside the patient.The procedure was completed with another jagwire guidewire.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
A visual evaluation of the returned guidewire revealed that the tip of the guidewire was peeled, exposing the corewire.The tip of the metal corewire was not exposed.There was no evidence of corewire fracture.Sanding marks were found on the corewire.Evidence of adhesive remnants were found which indicate that the pebax tip was correctly adhered during the manufacturing process.The complaint is not consistent with the returned guidewire since the tip was peeled and the tip of the metal corewire was not exposed.Based on all gathered information, the investigation fails to determine a definite root cause.There is an investigation in place to address distal tip related issues.A dhr (device history record) review was performed and no deviation was found.A labeling review was performed and there is no evidence that the device was not used in accordance with the labeling.
 
Event Description
It was reported to boston scientific corporation that a jagwire guidewire was used during an endoscopic retrograde cholangiopancreatography procedure performed on (b)(6) 2015.According to the complainant, during introduction, the hydrophilic tip became detached, exposing the tip of the metal corewire.No part of the guidewire detached inside the patient.The procedure was completed with another jagwire guidewire.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAGWIRE¿
Type of Device
ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS  
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key5317312
MDR Text Key34076181
Report Number3005099803-2015-03597
Device Sequence Number1
Product Code OCY
Combination Product (y/n)N
Reporter Country CodeCN
PMA/PMN Number
K141820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/27/2018
Device Model NumberM00556581
Device Catalogue Number5658
Device Lot Number17836646
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/02/2015
Initial Date FDA Received12/22/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/17/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age86 YR
Patient Weight55
-
-