Model Number EGIAADAPTXL |
Device Problem
Unintended Arm Motion (1033)
|
Patient Problem
Not Applicable (3189)
|
Event Date 12/01/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
According to the reporter, during a demonstration, when the reload is attached, it will articulate on its own, and the jaws will open on their own.When the reload moves back to neutral position, it is no longer recognized.
|
|
Manufacturer Narrative
|
Post market vigilance (pmv) led an evaluation of one device.This evaluation was based on a technical review of all data received from the site, a pmv review of manufacturing records, a pmv review of complaint trends, and an evaluation of the returned device by both pmv and engineering.The i investigation was able to replicate the reported conditions.Further investigation found cracked solder joints between the reload detect switch and the mma board.This can lead to intermittent occurrences of the reload not recognized and uncontrolled articulation conditions.Intermittent reload recognition due to a malfunctioning switch is the result of several variables including: improper solder operation at the vendor location, improper assembly of the sealed switch to the mma board at the vendor location, and/or improper soldering during assembly.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|