• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN, FORMERLY US SURGICAL A DIVISON ENDO GIA ADAPTER XL; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN, FORMERLY US SURGICAL A DIVISON ENDO GIA ADAPTER XL; STAPLE, IMPLANTABLE Back to Search Results
Model Number EGIAADAPTXL
Device Problem Unintended Arm Motion (1033)
Patient Problem Not Applicable (3189)
Event Date 12/01/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
According to the reporter, during a demonstration, when the reload is attached, it will articulate on its own, and the jaws will open on their own.When the reload moves back to neutral position, it is no longer recognized.
 
Manufacturer Narrative
Post market vigilance (pmv) led an evaluation of one device.This evaluation was based on a technical review of all data received from the site, a pmv review of manufacturing records, a pmv review of complaint trends, and an evaluation of the returned device by both pmv and engineering.The i investigation was able to replicate the reported conditions.Further investigation found cracked solder joints between the reload detect switch and the mma board.This can lead to intermittent occurrences of the reload not recognized and uncontrolled articulation conditions.Intermittent reload recognition due to a malfunctioning switch is the result of several variables including: improper solder operation at the vendor location, improper assembly of the sealed switch to the mma board at the vendor location, and/or improper soldering during assembly.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA ADAPTER XL
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN, FORMERLY US SURGICAL A DIVISON
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN, FORMERLY US SURGICAL A DIVISON
60 middletown ave
north haven CT 06473
Manufacturer Contact
sharon murphy
60 middletown ave
north haven, CT 06473
2034925267
MDR Report Key5319901
MDR Text Key34882502
Report Number1219930-2015-01087
Device Sequence Number1
Product Code GDW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121510
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/01/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberEGIAADAPTXL
Device Catalogue NumberEGIAADAPTXL
Device Lot NumberN5K0062LX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/01/2015
Initial Date FDA Received12/22/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/29/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-