• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPACELABS HEALTHCARE INC. SPACELABS TELEMETRY RECIEVER HOUSING; TELEMETRY HOUSING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPACELABS HEALTHCARE INC. SPACELABS TELEMETRY RECIEVER HOUSING; TELEMETRY HOUSING Back to Search Results
Model Number 90479
Device Problem Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/23/2015
Event Type  malfunction  
Manufacturer Narrative
The product was received at spacelabs¿ equipment service center for repair.The reported problem was verified.Battery assembly (b)(4) and pcba (b)(4) power supply were not functioning; therefore the parts were replaced.The repaired unit passed all functional tests and was returned to the customer.Lack of an ecg trace was the warning sign that the customer noticed and that prompted the action of removing the product from use.This report is complete and this particular issue is considered closed.Spacelabs monitors its complaints for similar product issues to determine need for additional investigation.
 
Event Description
Spacelabs received telemetry module housing model 90479 for service repair with customer complaint of no power.The customer removed the product from service on (b)(6) 2015.No injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPACELABS TELEMETRY RECIEVER HOUSING
Type of Device
TELEMETRY HOUSING
Manufacturer (Section D)
SPACELABS HEALTHCARE INC.
35301 se center st.
snoqualmie WA 98065
Manufacturer Contact
john zeng
35301 se center st.
snoqualmie, WA 98065
4253635915
MDR Report Key5320680
MDR Text Key34922423
Report Number3010157426-2015-00185
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K925510
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 12/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number90479
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/23/2015
Initial Date FDA Received12/22/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-