The product was received at spacelabs¿ equipment service center for repair.The reported problem was verified.Battery assembly (b)(4) and pcba (b)(4) power supply were not functioning; therefore the parts were replaced.The repaired unit passed all functional tests and was returned to the customer.Lack of an ecg trace was the warning sign that the customer noticed and that prompted the action of removing the product from use.This report is complete and this particular issue is considered closed.Spacelabs monitors its complaints for similar product issues to determine need for additional investigation.
|