MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SUMMIT POR TAPER SZ1 STD OFF; HIP FEMORAL STEM/SLEEVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SUMMIT POR TAPER SZ1 STD OFF; HIP FEMORAL STEM/SLEEVE Back to Search Results
Catalog Number 157001070
Device Problems Corroded (1131); Adverse Event Without Identified Device or Use Problem (2993); Appropriate Term/Code Not Available (3191)
Patient Problems Host-Tissue Reaction (1297); Ossification (1428); Pain (1994); Synovitis (2094); Joint Swelling (2356); Osteolysis (2377); Limited Mobility Of The Implanted Joint (2671); No Code Available (3191)
Event Date 10/07/2014
Event Type  Injury  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is the subject of litigation or a legal claim and complete product detail is not available at this time.A follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Event Description
Patient was revised.No reason for revision will be provided.Update rec'd (b)(6) 2015: litigation papers allege that corrosion and friction wear is believed to have caused amounts of toxic cobalt-chromium metal debris to be released into the patient's tissue surrounding the implant.The patient began to experience pain and difficulty with the implant following the surgery.
 
Manufacturer Narrative
No device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review was not possible because the required lot codes were not provided.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Manufacturer Narrative
(b)(4).This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Event Description
Update 4/13/16- pfs and medical records received.Pfs and medical records reviewed for mdr reportability.Pfs reported difficulty walking long distances/rising from knees/kneeling/sitting or standing for long periods of time and walks with a limp.Radiograph reportedly showed osteolysis and heterotopic ossification.Medical records reported moderate swelling and noise in the joint.Revision surgical report noted adverse local tissue reaction, large pseudotumor, significant bone loss of femur including greater trochanter and surgeon was unable to remove liner from the cup and had to revise both.There was no report of metal debris or corrosion.Metal ion lab results greater than 7 parts per billion.Pathology results reported synovitis.Cup added for malfunction and part/lot updated.The complaint was updated on: may 4, 2016.
 
Manufacturer Narrative
Additional narrative: no device associated with this report was received for examination.A complaint database search did not find any additional related reports against the acetabular cup, liner and femoral stem product code and lot number combination(s).The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Event Description
Ppf alleges metallosis.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Udi: (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: product complaint (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.Per (b)(4) a medical record review is not required for this complaint record.Investigational inputs were requested as indicated per internal procedures for this failure mode.The complaint information provided has been reviewed for complaint coding, medical device reporting, and other data required by the complaint system.Follow-up for additional event information, if applicable, was conducted utilizing work instruction wi-7915 appendix a.Without the physical complaint sample(s) associated with this report, it was not possible to determine if the device(s) failed to meet specification(s) at the time it was released for distribution.The device(s) associated with this event were used in the treatment of the patient as prescribed by the presiding surgeon.From the event information received, it was not possible to determine the relationship of the device to the reported event.No information received with this individual complaint indicated that a broader investigation or corrective action was necessary.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT POR TAPER SZ1 STD OFF
Type of Device
HIP FEMORAL STEM/SLEEVE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key5321553
MDR Text Key34197891
Report Number1818910-2015-38154
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
PMA/PMN Number
PK001991
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 12/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Catalogue Number157001070
Device Lot NumberB61EA1000
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/16/2015
Initial Date FDA Received12/22/2015
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
04/11/2019
01/12/2020
02/17/2020
Supplement Dates FDA Received02/18/2016
05/05/2016
08/12/2016
05/08/2019
01/12/2020
02/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age47 YR
Patient Weight54
-
-