STAAR SURGICAL COMPANY ICL (IMPLANTABLE COLLAMER LENS); INTRAOCULAR LENS Back to Search Results
Model Number MICL13.2
Device Problems Difficult to Advance (2920); Torn Material (3024)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/16/2015
Event Type  malfunction  
Manufacturer Narrative
Method: (process evaluation): work order search results: (evaluation result): a lens work order search was performed and another similar complaint was found within the work order.Conclusion: (unable to confirm complaint): based on the complaint history and work order search, a specific root cause of the event could not be determined.(b)(4).
 
Event Description
The reporter stated the surgeon attempted to insert a 13.2mm micl13.2 implantable collamer lens, -14.0 diopter but the lens became stuck in the cartridge and tore.The reporter stated it was unknown if there was any patient contact.Additional information has been requested but none has been forthcoming.If additional information is received, a supplemental medwatch report will be submitted.
 
Manufacturer Narrative
Method: device history record review, medical review.Results: a review of the device history record was performed and nothing was found in the manufacturing, inspection and packaging process records to suggest a contributory factor to the complaint.Medical review - reportedly, micl (13.2 mm) was stuck in the injector and was subsequently torn off during insertion requiring intraoperative exchange.No reported incision enlargement or any other tissue damage.According to report , by an office manager, dated on (b)(6) 2015 there was no confirmation if the lens had a patient contact.No reported postoperative sequelae.Visual inspection of the returned product found two lenses stuck together in one vial.The lenses were returned dry and could not be evaluated due to the returned condition.Conclusion: based on the complaint history, work order search, device history record review, medical review and the returned product, a specific root cause of the event could not be determined.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICL (IMPLANTABLE COLLAMER LENS)
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
althea watson
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key5323172
MDR Text Key34270439
Report Number2023826-2015-01722
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/01/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date06/30/2017
Device Model NumberMICL13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/01/2015
Initial Date FDA Received12/23/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/19/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTRIDGE MODEL AND LOT NUMBER UNK
Patient Age42 YR
-
-