• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA PRESSURE WIRE AERIS G8; TRANSDUCER, PRESSURE,CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA PRESSURE WIRE AERIS G8; TRANSDUCER, PRESSURE,CATHETER TIP Back to Search Results
Model Number C12058
Device Problem Use of Device Problem (1670)
Patient Problem No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
(b)(4).The results of the investigation are inconclusive since the device was not returned for analysis.A review of the device history record was not possible since the batch number was unavailable.Based on the information received, the cause of the reported incident could not be conclusively determined.The pressure wire instructions for use (ifu) states that torqueing the pressure wire against resistance or repeated attempts to cross a total vessel occlusion may cause damage and/or fracture, which may lead to a portion of pressure wire separating from the tip.The pressurewire instructions for use (ifu) states that the user should advance or withdraw the pressurewire slowly and never push, withdraw or torque the pressure wire if it meets resistance.
 
Event Description
When advancing the pressure wire aeris was to the lesion friction was noted due to calcification in the vessel.Following ffr measurement, the pressure wire was removed at the end of the procedure, it was noted that the radiopaque tip was detached and remained in the coronary artery.A snare was used to remove the tip without success.Due to the calcification of the artery, it was stated the patient was a candidate for bypass and the detached portion of the pressure wire aeris will be recovered during surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE WIRE AERIS G8
Type of Device
TRANSDUCER, PRESSURE,CATHETER TIP
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key5323432
MDR Text Key34248924
Report Number3008452825-2015-00126
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K131452
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial
Report Date 12/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberC12058
Device Catalogue NumberC12058
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/09/2015
Initial Date FDA Received12/23/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-