U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON CONCHA NEPTUNE; HEATED HUMIDIFICATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON CONCHA NEPTUNE; HEATED HUMIDIFICATION SYSTEM Back to Search Results
Catalog Number 425-00
Device Problem Power Problem (3010)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/11/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the device was not returned at the time of this report.The device history record (dhr) review did not show issues related to the complaint.The device was manufactured on 07/11/2008.A document assessment (fmea-08-037) was conducted and no changes were required.The reported complaint cannot be confirmed since the device sample is not available to perform a proper investigation, determine the root cause and assign corrective actions.If the device sample becomes available at a later date, this complaint will be updated accordingly.
 
Event Description
The event is reported as: the unit is shutting down during use.No harm or injury to patient reported.
 
Manufacturer Narrative
Qn#(b)(4).Serial # was corrected to (b)(4).A device history record (dhr) review was performed on lot number 608070236, and there were no issues found that could relate to the reported complaint.The sample was returned for evaluation.A visual exam was performed and no defects were observed.Functional testing was also performed and the unit was connected to 110vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test, the neptune displayed the correct temperatures and properly alarmed in the high temperature scenario.The unit was prepared for the functional bench test where water, an adult breathing circuit (880-36kit), 10 lpm of air pressure, and a low compliance column is connected to the unit for a real time operational scenario.The unit successfully negotiated all pre-operational self-tests again and was functioning real time.The unit functioned without interruption for two (2) hours.Based on the investigation performed, the reported complaint could not be confirmed.There were no issues found with the returned device.The unit functioned as intended.
 
Event Description
The event is reported as: the unit is shutting down during use.No harm or injury to patient reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHA NEPTUNE
Type of Device
HEATED HUMIDIFICATION SYSTEM
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key5323682
MDR Text Key34267659
Report Number3003898360-2015-00922
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/11/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number425-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/11/2015
Initial Date FDA Received12/23/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/17/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/11/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-