• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. DELTA HIP; HEAD, FEMORAL, CERAMIC, BILOX DELTA, 36MM NEUTRAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. DELTA HIP; HEAD, FEMORAL, CERAMIC, BILOX DELTA, 36MM NEUTRAL Back to Search Results
Catalog Number 400-03-362
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Code Available (3191)
Event Date 12/08/2015
Event Type  Injury  
Event Description
Revision surgery - the patient dislocated due to non compliance.
 
Manufacturer Narrative
The reason for this revision surgery was dislocation.The length of the in-vivo service was 21 days.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing history of this part.All parts were found to meet design and manufacturing specifications.No non-conforming material reports (ncmr's) were associated with this part.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed to be non-product related.The surgeon reported no issues associated with the explanted product and provided no information that definitively described the root cause or reason of the joint dislocation.The scope of this investigation is limited without having the parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTA HIP
Type of Device
HEAD, FEMORAL, CERAMIC, BILOX DELTA, 36MM NEUTRAL
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
teffany hutto
9800 metric blvd.
austin, TX 78758-5445
5128346255
MDR Report Key5324444
MDR Text Key34276484
Report Number1644408-2015-00766
Device Sequence Number1
Product Code LZO
UDI-Device Identifier00888912076135
UDI-Public(01)00888912076135
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082844
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/30/2020
Device Catalogue Number400-03-362
Device Lot Number866B1158
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/08/2015
Initial Date FDA Received12/23/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/07/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/28/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age72 YR
-
-