EDWARDS LIFESCIENCES DR DISPOSABLE PRESSURE TRANSDUCER; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number T430023M
Device Problem Difficult To Position (1467)
Patient Problem Skin Inflammation (2443)
Event Date 01/01/2015
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for evaluation; it was discarded at the hospital.Without return of the unit it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.It is not known if some procedural factors may have contributed to the event.No corrective actions will be taken at this time.Lot number was not provided; therefore review of the manufacturing records could not be completed.
 
Event Description
As reported, during use in patient in two occasions, the stand-alone stopcock of the dpt (disposable pressure transducer) was placed on the top of the patient's hand when connecting it to an arterial line, causing decubitus (pressure wound) on the patient.A picture of these cases could not be provided.It should be noticed that if the nurse totally adherent the connection to the patients wrist position, the risk of decubitus would only increase if the patient has edema.Furthermore, once the patient has developed a decubitus due to this connection there is an increased risk of infection and septic.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE PRESSURE TRANSDUCER
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR 
Manufacturer (Section G)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR  
Manufacturer Contact
lynn selawski
1 edwards way
irvine, CA 92614
9497564386
MDR Report Key5325042
MDR Text Key34290640
Report Number2015691-2015-03519
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeDA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial
Report Date 12/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberT430023M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/02/2015
Initial Date FDA Received12/23/2015
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-