SORIN GROUP DEUTSCHLAND GMBH S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 11/28/2015
Event Type  No Answer Provided  
Manufacturer Narrative
There was no patient involvement.Device has not been returned to sorin group (b)(4).Sorin group (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the touch screen of the s5 roller pump was unresponsive during priming.There was no patient involvement.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that the touch screen of the s5 roller pump was unresponsive during priming.There was no patient involvement.
 
Manufacturer Narrative
Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).Livanova replaced the touchscreen and the pump, and the unit has been returned to service.The touch screen was not returned for evaluation.Photos taken showed liquid penetrated into the flat cable and oxidation was visible.Root-cause was determined to be dried fluid present on the capton tail of defective touchscreen.A review of the dhr did not identify any deviations or nonconformities relevant to the reported issue.This is a known issue.Livanova (b)(4) has initiated capa (b)(4) for this type of issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND GMBH
lindberghstr. 25
munich,
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key5325887
MDR Text Key34302582
Report Number9611109-2015-00619
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/29/2015
Initial Date FDA Received12/23/2015
Supplement Dates Manufacturer Received09/21/2017
Supplement Dates FDA Received09/22/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/26/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-