Model Number 10-80-00 |
Device Problem
No Display/Image (1183)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/28/2015 |
Event Type
No Answer Provided
|
Manufacturer Narrative
|
There was no patient involvement.Device has not been returned to sorin group (b)(4).Sorin group (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the touch screen of the s5 roller pump was unresponsive during priming.There was no patient involvement.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
|
|
Event Description
|
Sorin group (b)(4) received a report that the touch screen of the s5 roller pump was unresponsive during priming.There was no patient involvement.
|
|
Manufacturer Narrative
|
Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).Livanova replaced the touchscreen and the pump, and the unit has been returned to service.The touch screen was not returned for evaluation.Photos taken showed liquid penetrated into the flat cable and oxidation was visible.Root-cause was determined to be dried fluid present on the capton tail of defective touchscreen.A review of the dhr did not identify any deviations or nonconformities relevant to the reported issue.This is a known issue.Livanova (b)(4) has initiated capa (b)(4) for this type of issue.
|
|