U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY ICL (IMPLANTABLE COLLAMER LENS); INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY ICL (IMPLANTABLE COLLAMER LENS); INTRAOCULAR LENS Back to Search Results
Model Number VTICMO12.1
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Code Available (3191)
Event Date 07/22/2015
Event Type  Injury  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.(b)(4).Method: work order search.Results: a lens work order search was performed and no similar complaints were found within the work order.Conclusion: based on the complaint history and work order search, a specific root cause of the event could not be determined.(b)(4).
 
Event Description
The reporter indicated the surgeon implanted a 12.1mm vticmo12.1 implantable collamer lens, -9.5/+1.5/x057 diopter, in the patient's left eye (os) on (b)(6) 2015.The lens was explanted on (b)(6) 2015 due to a low vault.The lens was exchanged for a longer lens and the problem was resolved.The patient's post-op best-corrected visual acuity was 20/25.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICL (IMPLANTABLE COLLAMER LENS)
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
althea watson
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key5325950
MDR Text Key34301271
Report Number2023826-2015-01709
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial
Report Date 11/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date04/30/2018
Device Model NumberVTICMO12.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/16/2015
Initial Date FDA Received12/24/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/21/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTRIDGE MODEL AND LOT NUMBER UNK
Patient Outcome(s) Required Intervention;
Patient Age35 YR
-
-