The customer reported that they received a transport error on their mpa pre-analytics system.It was determined that there were 2 racks, containing ten samples, which did not get centrifuged.These racks were sent to 2 cobas 8000 analyzer lines, where they were processed.Of the ten samples, seven had erroneous results reported outside of the laboratory for creatinine plus ver.2 (crea), ion selective electrode (ise) potassium - k, magnesium gen.2 (mg), and glucose (glc).Refer to the attachment for results from all seven affected samples.The initial results were from the racks that were not centrifuged.The samples were then centrifuged and repeated.The repeat results were believed to be correct and corrected reports were issued.The patients were not adversely affected.The crea reagent lot number was 617455 and the mg reagent lot number was 618718.The crea and mg reagent expiration dates were asked for, but not provided.The ise potassium electrode and glc reagent lot numbers and expiration dates were asked for, but not provided.The customer found loose labels on the sample tubes and found part of a label on the belt of the mpa system.The label was removed from the belt.The customer re-set the mpa system and no further issues occurred.The customer declined a service visit.
|