It was reported that the patient underwent generator and lead explant surgery due to low battery and lead pulling sensation on (b)(6) 2015.The explanted generator and lead were received by the manufacturer for analysis.Analysis of the generator in the analysis lab concluded that no abnormal performance or any other type of adverse condition was found, other than normal battery depletion.Analysis of the lead, on the other hand, showed pitting or electro-etching conditions have occurred on the unmarked connector pin surface.Abraded openings and fluid leaks were observed in the outer silicon tubing.Note that since a portion of the lead (including the electrode array) was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.Other than the above mentioned observations and typical wear and explant related observations, no other anomalies were identified in the returned lead portion.
|