ABBOTT MANUFACTURING INC ACCELERATOR APS CENTRIFUGE MODULE; CENTRIFUGE MODULE FOR AUTOMATED SAMPLE HANDLING AND PROCESSING SYSTEM Back to Search Results
Catalog Number 07L02-02
Device Problem Charred (1086)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.(b)(4).An abbott field service representative (fsr) found a burned r100 component on the main electronics board, part number (b)(4).The board was replaced to resolve the issue.The accelerator aps operations manual and the centrifuge operations manual were reviewed.Troubleshooting and safety information is provided.Customer complaint data was reviewed and an increase in failures for the main electronics board, part (b)(4) triggered an alert which suggests an adverse trend.The failure prevents the centrifuge from performing as expected.Based on this information a product deficiency was identified and further investigation is being performed.A follow up report will be submitted at the completion of the investigation.Investigation ongoing.
 
Event Description
The customer described a burning odor coming from the aps centrifuge module (cm).The module was powered off and an abbott field service representative (fsr) was sent on site.After inspecting the module, the fsr found charring on the main electronics board.There were no injuries reported with no damage to the surrounding lab environment.There were no reports of any impact to patient managment.
 
Manufacturer Narrative
A product deficiency was identified due to an increase in complaints for failures of the r100 component of the main electronics board (parts 8-35003665-01 and 8-206623-01).The supplier, (b)(4), determined that the issue is from the r100 component of the hettich centrifuge board which was too small to withstand a current spike.Hettich released an enhanced main board revision 06 with a different r100 resistor.It is considered acceptable to continue using the automation system because the issue does not impact operator nor patient safety.The accelerator aps operations manual and the centrifuge operations manual were reviewed and show that troubleshooting and safety information is provided.There have been no reported injuries for this issue.If the electrical components of the main electronics board fail, it will shut down the centrifuge module.If the centrifuge shuts down, the operator will receive an error message.The frequency of this hazard was assessed and it was determined that though there was a trend in complaints, the overall frequency of low had not changed from the predicted frequency as outlined in the risk management file.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCELERATOR APS CENTRIFUGE MODULE
Type of Device
CENTRIFUGE MODULE FOR AUTOMATED SAMPLE HANDLING AND PROCESSING SYSTEM
Manufacturer (Section D)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
Manufacturer (Section G)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key5331594
MDR Text Key35035129
Report Number1628664-2015-00342
Device Sequence Number1
Product Code JQP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Medical Technologist
Type of Report Initial,Followup
Report Date 06/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07L02-02
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/22/2015
Initial Date FDA Received12/29/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-